Abstract
Abstract
Purpose
To assess multicenter two-year patency outcomes of endovascular arteriovenous fistulas (endoAVF) created with the WavelinQ device.
Materials and Methods
Patients who had fistulas created at three centers from January 2018 to December 2020 were included in this retrospective study. In total, 112 patients underwent endoAVF creation [40 females, 72 males; mean age 60 years (range 18–88)]. Data collected included patient demographics, location of fistula creation, interventions performed, and brachial artery flows pre- and post-creation. Two-year cumulative patency, functional patency, and primary patency were assessed with Kaplan–Meier methodology. Factors affecting patency and maturation were examined using the Cox proportional hazards model.
Results
Technical success defined as angiographically successful endoAVF creation was 97.3% (109/112). In 11 patients the fistula did not mature for dialysis use. For 98 patients (87%) with endoAVF maturation, 12- and 24-month cumulative patency was 94.3% and 91.7%. Functional patency (two-needle cannulation) at 12 and 24 months was 95.7% and 92.7%, respectively. Median maturation time is 95 days (IQR 51–231 days). Male gender and brachial vein coiling at the time of endoAVF creation were predictive of maturation. There were 34 censored events (four patients undergoing renal transplantation; 30 patients deceased). Number of reinterventions per patient year was 0.73 where 43 were maturation procedures and 101 were maintenance procedures. One Grade 3 complication occurred of arterial access puncture site pseudoaneurysm.
Conclusion
A high two-year functional and cumulative patency following endoAVF creation with the WavelinQ device was observed in this multicenter real-world experience Level of Evidence: 3
Level of Evidence III
Graphical Abstract
Publisher
Springer Science and Business Media LLC
Cited by
1 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献