In retrospective clinical studies digital variance angiography (DVA) provided higher contrast-to-noise ratio and better image quality than digital subtraction angiography (DSA). Our aim was to verify the clinical usefulness and benefits of DVA in carbon dioxide (CO2)-assisted lower limb interventions.
Materials and Methods
A workstation running the DVA software was integrated into a Siemens Artis Zee with Pure angiography system, and this new image processing technology was used in four patients (3 male, 1 female, age: 76.2 ± 4.2 years) with peripheral artery disease (PAD, Rutherford 2–3) and impaired renal function (average eGFR 25.5 ± 11.2 ml/min/1.73 m2). The DSA and DVA images of 46 CO2-assisted runs were visually evaluated by five experts in single-image evaluation using a 5-grade Likert scale and in paired comparisons.
Results
DVA images received significantly higher score (3.84 ± 0.10) than DSA images (3.31 ± 0.10, p < 0.001). Raters preferred DVA images in terms of diagnostic value and usefulness for therapeutic decisions in 85.2% and 83.9% of all comparisons, respectively. These benefits were achieved at lower frame rates (1–3 FPS) than usually recommended for CO2 angiography (4–6 FPS). No adverse events were recorded during or after the procedures.
Conclusions
Our initial experience shows that DVA might facilitate the correct diagnostic and therapeutic decisions, and potentially help to reduce radiation exposure in lower limb CO2 angiography. Although the dose management capabilities of DVA have to be validated in further clinical studies, this technology might be a useful new tool in the operating room and contributes to the safety and efficacy of CO2-enhanced endovascular interventions.
Level of Evidence
Level IV.