Purpose
Prognosis of patients presenting with INTERMACS 1 critical cardiogenic shock is generally poor. The aim of our study was to investigate the results of CentriMag™ extracorporeal short-term mechanical circulatory support as a bridge to decision in patients presenting with critical cardiogenic shock in our unit.
Methods
We retrospectively analysed 63 consecutive patients from January 2005 to June 2017, who were treated with a CentriMag™ device at our institution as a bridge to decision. Patients requiring extracorporeal support for post-cardiotomy shock and for primary graft dysfunction after heart transplantation were excluded.
Results
Patients’ median age was 44 years (IQR 31–52, range 15.4–62.0) and 42 (67%) were male. Primary diagnosis at presentation was ischaemic cardiomyopathy (n = 24; 38.1%), viral myocarditis (n = 19; 30.2%), idiopathic dilated cardiomyopathy (n = 8; 12.7%), and others (n = 12; 19%). The median duration of support was 25 (IQR 9.5–56) days. A total of 7 (11%) patients were supported with peripheral veno-arterial (VA) extra corporeal membrane oxygenation (ECMO), 6 (9%) with central VA ECMO, 8 (13%) with left ventricular assist device (LVAD), 17 (27%) with biventricular assist device (BiVAD), and 25 (40%) with ECMO and then converted to BiVAD. Overall, 22 (34.9%) patients died while on CentriMag™ mechanical circulatory support. Complications included bleeding requiring reoperation/intervention in 24 (38%), renal failure requiring dialysis in 29 (46%), bacterial infections in 23 (37%), fungal infections in 15 (24%), critical limb ischaemia in 6 (10%), and stroke in 8 (13%). The overall survival to successful explant from CentriMag™ was 65.1% (n = 41) and survival to hospital discharge was 58.7% (n = 37). Of these, 10 (16%) had cardiac recovery and were successfully explanted, 20 (32%) were bridged to heart transplantation, 11 (17%) were bridged to long-term left ventricular assist device, 3 (4.7%) were later on transplanted, and 1 (1.6%) recovered to decommissioning. The 1-, 5-, and 10-year survival rates were 55%, 46%, and 23% respectively.
Conclusion
Our results demonstrate an excellent outcome with the use of the CentriMag™ device in this seriously ill population. Despite requiring multiple procedures, over 58% of patients were discharged from hospital with 5-year survival of 46%.
Publisher
Springer Science and Business Media LLC
Subject
Cardiology and Cardiovascular Medicine,Pulmonary and Respiratory Medicine,Surgery
Reference14 articles.
1. UK annual report on mechanical circulatory support related to heart transplantation - report for 2018/2019 [Internet]. 2019. Available from: https://nhsbtdbe.blob.core.windows.net/umbraco-assets-corp/16956/nhsbt-annual-report-on-mcs-201819.pdf. Accessed 7 Apr 2020
2. Tongers J, Sieweke J-T, Kühn C, et al. Early escalation of mechanical circulatory support stabilizes and potentially rescues patients in refractory cardiogenic shock. Circ Heart Fail. 2020;13:e005853.
3. Nersesian G, Hennig F, Müller M, et al. Temporary mechanical circulatory support for refractory heart failure: the German Heart Center Berlin experience. Ann Cardiothorac Surg. 2019;8:76–83.
4. Mahboub-Ahari A, Heidari F, Sadeghi-Ghyassi F, Asadi M. A systematic review of effectiveness and economic evaluation of Cardiohelp and portable devices for extracorporeal membrane oxygenation (ECMO). J Artif Organs. 2019;22:6–13.
5. John R, Long JW, Massey HT, et al. Outcomes of a multicenter trial of the Levitronix CentriMag ventricular assist system for short-term circulatory support. J Thorac Cardiovasc Surg. 2011;141:932–9.
Cited by
13 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献