Author:
Razzetta Chiara,Candiani Valentina,Crocco Marco,Benvenuto Federico
Abstract
AbstractIn ultrasound imaging, state-of-the-art methods for transmit beam pattern (TBP) optimization have well-known drawbacks like non-uniform beam width over depth, significant side lobes, and quick energy drop-out beyond the focal region. To overcome these limitations, we develop a TBP optimization approach by focusing on its narrowband approximation and considering transmit delays as free variables instead of linked to specific focal depths. We formulate a non-linear Least Squares problem to minimize the difference between the TBP corresponding to a set of delays and the desired one, modeled as a 2D rectangle elongated along the beam axis. Three metrics are defined to quantitatively evaluate the results. Results obtained by synthetic simulations show that the main lobe width is considerably more intense ($$+31.5\%$$
+
31.5
%
on average) and uniform ($$+28.7\%$$
+
28.7
%
on average) over the whole depth range compared to classical focalized Beam Patterns. The application of the method to elastography shows improvements in the ultrasound energy concentration along a desired axis.
Publisher
Springer Science and Business Media LLC