Abstract
Abstract
Background
Hyperthermia treatment planning can be supportive to ensure treatment quality, provided reliable prediction of the heating characteristics (i.e., focus size and effects of phase-amplitude and frequency steering) of the device concerned is possible. This study validates the predictions made by the treatment planning system Plan2Heat for various clinically used phased-array systems.
Methods
The evaluated heating systems were AMC-2, AMC-4/ALBA-4D (Med-Logix srl, Rome, Italy), BSD Sigma-30, and Sigma-60 (Pyrexar Medical, Salt Lake City, UT, USA). Plan2Heat was used for specific absorption rate (SAR) simulations in phantoms representing measurement set-ups reported in the literature. SAR profiles from published measurement data based on E‑field or temperature rise were used to compare the device-specific heating characteristics predicted by Plan2Heat.
Results
Plan2Heat is able to predict the correct location and size of the SAR focus, as determined by phase-amplitude settings and operating frequency. Measured effects of phase-amplitude steering on focus shifts (i.e., local SAR minima or maxima) were also correctly reflected in treatment planning predictions. Deviations between measurements and simulations were typically < 10–20%, which is within the range of experimental uncertainty for such phased-array measurements.
Conclusion
Plan2Heat is capable of adequately predicting the heating characteristics of the AMC‑2, AMC-4/ALBA-4D, BSD Sigma-30, and Sigma-60 phased-array systems routinely used in clinical hyperthermia.
Publisher
Springer Science and Business Media LLC
Reference78 articles.
1. van der Zee J, González D, van Rhoon GC, van Dijk JDP, van Putten WLJ, Hart AA (2000) Comparison of radiotherapy alone with radiotherapy plus hyperthermia in locally advanced pelvic tumours: a prospective, randomised, multicentre trial. Dutch deep hyperthermia group. Lancet 355:1119–1125
2. Kok HP, Cressman ENK, Ceelen W, Brace CL, Ivkov R, Grull H, Ter Haar G, Wust P, Crezee J (2020) Heating technology for malignant tumors: a review. Int J Hyperthermia 37:711–741
3. Stauffer PR (2005) Evolving technology for thermal therapy of cancer. Int J Hyperthermia 21:731–744
4. Wust P, Seebass M, Nadobny J, Deuflhard P, Monich G, Felix R (1996) Simulation studies promote technological development of radiofrequency phased array hyperthermia. Int J Hyperthermia 12:477–494
5. Turner PF, Tumeh A, Schaefermeyer T (1989) BSD-2000 approach for deep local and regional hyperthermia: physics and technology. Strahlenther. Onkol 165:738–741