Pressure-adjusted venting eliminates start-up delays and compensates for vertical position of syringe infusion pumps used for microinfusion

Author:

Weiss Markus,Wendel-Garcia Pedro David,Cannizzaro Vincenzo,Grass Beate,Buehler Philipp Karl,Kleine-Brueggeney Maren

Abstract

AbstractMicroinfusions are commonly used for the administration of catecholamines, but start-up delays pose a problem for reliable and timely drug delivery. Recent findings show that venting of the syringe infusion pump with draining of fluid to ambient pressure before directing the flow towards the central venous catheter does not counteract start-up delays. With the aim to reduce start-up delays, this study compared fluid delivery during start-up of syringe infusion pumps without venting, with ambient pressure venting, and with central venous pressure (CVP)-adjusted venting. Start-up fluid delivery from syringe pumps using a microinfusion of 1 mL/h was assessed by means of liquid flow measurement at 10, 60, 180 and 360 s after opening the stopcock and starting the pump. Assessments were performed using no venting, ambient pressure venting or CVP-adjusted venting, with the pump placed either at zero, − 43 cm or + 43 cm level and exposed to a simulated CVP of 10 mmHg. Measured fluid delivery was closest to the calculated fluid delivery for CVP-adjusted venting (87% to 100% at the different timepoints). The largest deviations were found for ambient pressure venting (− 1151% to + 82%). At 360 s after start-up 72% to 92% of expected fluid volumes were delivered without venting, 46% to 82% with ambient pressure venting and 96% to 99% with CVP-adjusted venting. CVP-adjusted venting demonstrated consistent results across vertical pump placements (p = 0.485), whereas the other methods had significant variances (p < 0.001 for both). In conclusion, CVP-adjusted venting effectively eliminates imprecise drug delivery and start-up delays when using microinfusions.

Funder

University of Zurich

Publisher

Springer Science and Business Media LLC

Subject

Anesthesiology and Pain Medicine,Critical Care and Intensive Care Medicine,Health Informatics,Anesthesiology and Pain Medicine,Critical Care and Intensive Care Medicine,Health Informatics

Cited by 1 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3