Evaluation of the accuracy of oscillometric non-invasive blood pressure measurement at the ankle in children during general anesthesia

Author:

Gamal Reham M,Mostafa Maha,Hasanin Ahmed M,Khedr Suzan Adlan,Abdelgalil Ahmed Salah,Elshal Mamdouh Mahmoud

Abstract

AbstractThis study aimed to evaluate the accuracy of oscillometric blood pressure measurement at the ankle in children using invasive blood pressure as reference standard. This prospective observational study included children undergoing noncardiac surgery. Paired radial invasive and ankle non-invasive blood pressure measurements were obtained. Delta blood pressure was calculated as the difference between two consecutive readings. The primary outcome was the mean bias and agreement between the two methods using the Bland-Altman analysis. The ISO standard was fulfilled if the mean bias between the two methods was ≤ 5 ± 8 mmHg. Other outcomes included the trending ability of ankle blood pressure using the four-quadrant plot and the accuracy of ankle measurement to detect hypotension using area under receiver operating characteristic curve (AUC) analysis. We analyzed 683 paired readings from 86 children. The mean bias between the two methods for systolic, diastolic, and mean blood pressure (SBP, DBP, MAP) was − 7.2 ± 10.7, 4.5 ± 12.8, and − 1.8 ± 8.2 mmHg, respectively. The concordance rate of ankle blood pressure was 72%, 71%, and 77% for delta SBP, DBP and MAP, respectively. The AUC (95% confidence interval) for ankle MAP ability to detect hypotension was 0.91 (0.89–0.93) with negative predictive value of 100% at cut-off value ≤ 70 mmHg, We concluded that in pediatric population undergoing noncardiac surgery, ankle blood pressure was not interchangeable with the corresponding invasive readings with the ankle MAP having the least bias compared to SBP and DBP. An ankle MAP > 70 mmHg can exclude hypotension with negative predictive value of 100%.

Funder

Cairo University

Publisher

Springer Science and Business Media LLC

Subject

Anesthesiology and Pain Medicine,Critical Care and Intensive Care Medicine,Health Informatics,Anesthesiology and Pain Medicine,Critical Care and Intensive Care Medicine,Health Informatics

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3