Abstract
AbstractSpinal fixation and fusion are surgical procedures undertaken to restore stability in the spine and restrict painful or degenerative motion. Malpositioning of pedicle screws during these procedures can result in major neurological and vascular damage. Patient-specific surgical guides offer clear benefits, reducing malposition rates by up to 25%. However, they suffer from long lead times and the manufacturing process is dependent on third-party specialists. The development of a standard set of surgical guides may eliminate the issues with the manufacturing process. To evaluate the feasibility of this option, a statistical shape model (SSM) was created and used to analyse the morphological variations of the T4–T6 vertebrae in a population of 90 specimens from the Visible Korean Human dataset (50 females and 40 males). The first three principal components, representing 39.7% of the variance within the population, were analysed. The model showed high variability in the transverse process (~ 4 mm) and spinous process (~ 4 mm) and relatively low variation (< 1 mm) in the vertebral lamina. For a Korean population, a standardised set of surgical guides would likely need to align with the lamina where the variance in the population is lower. It is recommended that this standard set of surgical guides should accommodate pedicle screw diameters of 3.5–6 mm and transverse pedicle screw angles of 3.5°–12.4°.
Publisher
Springer Science and Business Media LLC
Subject
Mechanical Engineering,Modeling and Simulation,Biotechnology
Cited by
1 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献