Abstract
Abstract
Purpose of Review
The risk of cardiac implantable electronic device (CIED) interference from cell phones was previously thought to be low based on older studies. Current generation of smartphones have incorporated more magnets for optimization of wireless charging, attachment of accessories, and convenience functionalities. These magnets have the potential to cause CIEDs to inadvertently revert into magnet mode. The purpose of this review is to summarize recent findings on smartphones and their accessories causing interference on CIEDs.
Recent Findings
Recent reports have demonstrated that the iPhone 12 series and accessories have the capability to cause CIED magnetic interference.
Summary
Current generation of smartphones, smartwatches, wireless headphones, and accessories have the potential to cause CIEDs to revert into magnet mode in both in vivo and ex vivo experiments. The risk of a clinically significant event is unlikely as long as the Food and Drug Administration (FDA) recommendations are followed; keeping smartphones and accessories at least six inches away from CIEDs.
Publisher
Springer Science and Business Media LLC
Subject
Cardiology and Cardiovascular Medicine
Reference39 articles.
1. Demographics of mobile device ownership and adoption in the United States. Pew Research Center: Internet, Science & Tech. https://www.pewresearch.org/internet/fact-sheet/mobile/. Published April 7, 2021. Accessed 21 Oct 2021.
2. Mond HG, Proclemer A. The 11th world survey of cardiac pacing and implantable cardioverter-defibrillators: calendar year 2009—a World Society of Arrhythmia’s project. Pacing Clin Electrophysiol. 2011;34(8):1013–27.
3. Hayes DL, Wang PJ, Reynolds DW, Estes M, Griffith JL, Steffens RA, et al. Interference with cardiac pacemakers by cellular telephones. N Engl J Med. 1997;336(21):1473–9.
4. Barbaro V, Bartolini P, Donato A, Militello C, Altamura G, Ammirati F, et al. Do European GSM mobile cellular phones pose a potential risk to pacemaker patients? Pacing Clin Electrophysiol. 1995;18(6):1218–24.
5. Altamura G, Toscano S, Gentilucci G, Ammirati F, Castro A, Pandozi C, et al. Influence of digital and analogue cellular telephones on implanted pacemakers. Eur Heart J. 1997;18(10):1632–41.
Cited by
5 articles.
订阅此论文施引文献