Author:
Sakurai Kenji,Hosoya Hiromi,Kurihara Yoshitaka,Yamauchi Fumi,Suzuki Ayumi,Kurosawa Kaori,Saito Takeshi
Abstract
Abstract
Purpose
To assess the removal performance of low-volume post-hemodiafiltration (HDF) with Japanese hemodiafilters and the removal performance with 20 % reduction in the total dialysate flow rate (Qdtotal).
Methods
Subjects were 8 patients undergoing pre-HDF. Study 1: Post-HDF was performed at a blood flow rate (Qb) of 250 mL/min and a total volume of substitution fluid (Vs) of 12 L/session(s) for 4 hrs using Fineflux-210Seco (FIX), ABH-21PA (ABH), and NVF-21H (NVF). We assessed removal efficiency of small molecular solutes, low-molecular-weight-proteins and the amount of albumin loss. Study 2: Post-HDF was performed at Vs of 12 L/s under G-1, Qdtotal of 500 and Qb of 250 mL/min; G-2, Qdtotal of 400 and Qb of 250 mL/min; and G-3, Qdtotal of 400 and Qb of 300 mL/min. Removal efficiency was compared and analyzed between these conditions.
Results
Study 1: The results using FIX, ABH and NVF are shown in order. The Kt/V were 1.8, 1.9 and 1.8. The β2-Microglobulin (MG) removal rate (RR) (%) were 81.2, 83.1 and 82.8, and the α1-MG RR were 37.4, 40.2 and 38.5, respectively. Study 2: The results in G-1, 2 and 3 are shown in order. The Kt/V and the RR of small solutes, were significantly higher in G-3. The β2-MG RR (%) were 81.2, 80.1 and 81.0, and the α1-MG RR were 37.4, 37.5 and 38.0, respectively.
Conclusions
Low-volume post-HDF performed at Qb of 250 mL/min with Japanese high-performance hemodiafilters exhibited favorable removal efficiency for all solutes. Even with 20 % reduction in Qdtotal, the removal performance was also favorable.
Publisher
Springer Science and Business Media LLC
Subject
Cardiology and Cardiovascular Medicine,Biomedical Engineering,Biomaterials,Medicine (miscellaneous)
Cited by
3 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献