Review of Adverse Events Associated With False Glucose Readings Measured by GDH-PQQ–Based Glucose Test Strips in the Presence of Interfering Sugars

Author:

Frias Juan P.1,Lim Christine G.1,Ellison John M.1,Montandon Carol M.1

Affiliation:

1. From the Departments of Clinical, Medical, and Regulatory Affairs, LifeScan, Milpitas, California.

Abstract

OBJECTIVE To assess the implications of falsely elevated glucose readings measured with glucose dehydrogenase pyrroloquinolinequinone (GDH-PQQ) test strips. RESEARCH DESIGN AND METHODS We conducted a review of the Food and Drug Administration's Manufacturer and User Facility Device Experience database and medical literature for adverse events (AEs) associated with falsely elevated glucose readings with GDH-PQQ test strips in the presence of interfering sugars. RESULTS Eighty-two reports were identified: 16 (20%) were associated with death, 46 (56%) with severe hypoglycemia, and 12 (15%) with nonsevere hypoglycemia. In eight reports (10%), the AE was not described. Forty-two events (51%) occurred in the U.S. Although most events occurred in hospitalized patients, at least 14 (17%) occurred in outpatients. Agents most commonly associated with AEs were icodextrin-containing peritoneal dialysate and maltose-containing intravenous immune globulin. CONCLUSIONS GDH-PQQ test strips pose a safety risk to insulin-using patients treated with agents containing or metabolized to interfering sugars.

Publisher

American Diabetes Association

Subject

Advanced and Specialized Nursing,Endocrinology, Diabetes and Metabolism,Internal Medicine

Reference7 articles.

1. Food and Drug Administration: FDA public health notification: potentially fatal errors with GDH-P glucose monitoring technology [article online], 2009. Available from http://www.fda.gov/MedicalDevices/Safety/AlertsandNotices/PublicHealthNotifications/ucm176992.htm. Accessed 13 August 2009

2. Frequency of hypoglycemia and its significance in chronic kidney disease;Moen;Clin J Am Soc Nephrol,2009

3. Unlabeled uses of intravenous immune globulin;Leong;Am J Health-Sys Pharm,2008

4. Current usage of intravenous immune globulin and the rationale behind it: Massachusetts General Hospital data and a review of the literature;Darabi;Transfusion,2006

5. Limitations and strengths of spontaneous report data;Goldman;Clin Thera,1998

Cited by 59 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献