MD-Logic Overnight Control for 6 Weeks of Home Use in Patients With Type 1 Diabetes: Randomized Crossover Trial

Author:

Nimri Revital1,Muller Ido1,Atlas Eran1,Miller Shahar1,Fogel Aviel1,Bratina Natasa2,Kordonouri Olga3,Battelino Tadej24,Danne Thomas3,Phillip Moshe15

Affiliation:

1. The Jesse Z. and Sara Lea Shafer Institute for Endocrinology and Diabetes, National Center for Childhood Diabetes, Schneider Children’s Medical Center of Israel, Petah Tikva, Israel

2. Department of Pediatric Endocrinology, Diabetes and Metabolism, University Medical Center, University Children’s Hospital, Ljubljana, Slovenia

3. Diabetes Center for Children and Adolescents, AUF DER BULT, Kinder- und Jugendkrankenhaus, Hannover, Germany

4. Faculty of Medicine, University of Ljubljana, Ljubljana, Slovenia

5. Sackler Faculty of Medicine, Tel Aviv University, Tel Aviv, Israel

Abstract

OBJECTIVE We evaluated the effect of the MD-Logic system on overnight glycemic control at patients' homes. RESEARCH DESIGN AND METHODS Twenty-four patients (aged 12–43 years; average A1c 7.5 ± 0.8%, 58.1 ± 8.4 mmol/mol) were randomly assigned to participate in two overnight crossover periods, each including 6 weeks of consecutive nights: one under closed loop and the second under sensor-augmented pump (SAP) therapy at patients' homes in real-life conditions. The primary end point was time spent with sensor glucose levels below 70 mg/dL (3.9 mmol/L) overnight. RESULTS Closed-loop nights significantly reduced time spent in hypoglycemia (P = 0.02) and increased the percentage of time spent in the target range of 70–140 mg/dL (P = 0.003) compared with nights when the SAP therapy was used. The time spent in substantial hyperglycemia above 240 mg/dL was reduced by a median of 52.2% (interquartile range [IQR] 4.8, 72.9%; P = 0.001) under closed-loop control compared with SAP therapy. Overnight total insulin doses were lower in the closed-loop nights compared with the SAP nights (P = 0.04). The average daytime glucose levels after closed-loop operation were reduced by a median of 10.0 mg/dL (IQR −2.7, 19.2; P = 0.017) while lower total insulin doses were used (P = 0.038). No severe adverse events occurred during closed-loop control; there was a single event of severe hypoglycemia during a control night. CONCLUSIONS The long-term home use of automated overnight insulin delivery by the MD-Logic system was found to be a feasible, safe, and an effective tool to reduce nocturnal hypoglycemia and improve overnight glycemic control in subjects with type 1 diabetes.

Publisher

American Diabetes Association

Subject

Advanced and Specialized Nursing,Endocrinology, Diabetes and Metabolism,Internal Medicine