Closed-Loop Insulin Delivery Using a Subcutaneous Glucose Sensor and Intraperitoneal Insulin Delivery

Author:

Renard Eric1,Place Jerome1,Cantwell Martin2,Chevassus Hugues3,Palerm Cesar C.2

Affiliation:

1. Endocrinology Department, Le Centre Hospitalier Universitaire Montpellier and Unité Mixte de Recherche Centre National de la Recherche Scientifique 5232, University of Montpellier, Montpellier, France;

2. Medtronic Diabetes, Northridge, California;

3. INSERM Clinical Investigation Centre 001, Le Centre Hospitalier Universitaire Montpellier, Montpellier, France.

Abstract

OBJECTIVE Attempts to build an artificial pancreas by using subcutaneous insulin delivery from a portable pump guided by an subcutaneous glucose sensor have encountered delays and variability of insulin absorption. We tested closed-loop intraperitoneal insulin infusion from an implanted pump driven by an subcutaneous glucose sensor via a proportional-integral-derivative (PID) algorithm. RESEARCH DESIGN AND METHODS Two-day closed-loop therapy (except for a 15-min premeal manual bolus) was compared with a 1-day control phase with intraperitoneal open-loop insulin delivery, according to randomized order, in a hospital setting in eight type 1 diabetic patients treated by implanted pumps. The percentage of time spent with blood glucose in the 4.4–6.6 mmol/l range was the primary end point. RESULTS During the closed-loop phases, the mean ± SEM percentage of time spent with blood glucose in the 4.4–6.6 mmol/l range was significantly higher (39.1 ± 4.5 vs. 27.7 ± 6.2%, P = 0.05), and overall dispersion of blood glucose values was reduced among patients. Better closed-loop glucose control came from the time periods excluding the two early postprandial hours with a higher percentage of time in the 4.4–6.6 mmol/l range (46.3 ± 5.3 vs. 28.6 ± 7.4, P = 0.025) and lower mean blood glucose levels (6.9 ± 0.3 vs. 7.9 ± 0.6 mmol/l, P = 0.036). Time spent with blood glucose <3.3 mmol/l was low and similar for both investigational phases. CONCLUSIONS Our results demonstrate the feasibility of intraperitoneal insulin delivery for an artificial β-cell and support the need for further study. Moreover, according to a semiautomated mode, the features of the premeal bolus in terms of timing and amount warrant further research.

Publisher

American Diabetes Association

Subject

Advanced and Specialized Nursing,Endocrinology, Diabetes and Metabolism,Internal Medicine

Cited by 150 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

1. Wireless Battery-free and Fully Implantable Organ Interfaces;Chemical Reviews;2024-02-21

2. Controlling Human Physiology via Drugs and Devices;Reference Module in Materials Science and Materials Engineering;2024

3. Artificial Pancreas Blood Glucose-Insulin Regularization System Using PID Controller;2023 Innovations in Power and Advanced Computing Technologies (i-PACT);2023-12-08

4. Multivariable identification based MPC for closed-loop glucose regulation subject to individual variability;Computer Methods in Biomechanics and Biomedical Engineering;2023-11-20

5. Transdermal hormone delivery: Strategies, application and modality selection;Journal of Drug Delivery Science and Technology;2023-09