Abstract
OBJECTIVE
To determine the relationship between mean sensor glucose concentrations and hemoglobin A1c (HbA1c) values measured in the Diabetes Control and Complications Trial/Epidemiology of Diabetes Interventions and Complications laboratory at the University of Minnesota in a cohort of subjects with type 1 diabetes from the Juvenile Diabetes Research Foundation continuous glucose monitoring randomized trial.
RESEARCH DESIGN AND METHODS
Near-continuous glucose sensor data (≥4 days/week) were collected for 3 months before a central laboratory–measured HbA1c was performed for 252 subjects aged 8–74 years, the majority of whom had stable HbA1c values (77% within ±0.4% of the patient mean).
RESULTS
The slope (95% CI) for mean sensor glucose concentration (area under the curve) versus a centrally measured HbA1c was 24.4 mg/dL (22.0–26.7) for each 1% change in HbA1c, with an intercept of −16.2 mg/dL (−32.9 to 0.6). Although the slope did not vary with age or sex, there was substantial individual variability, with mean sensor glucose concentrations ranging from 128 to 187 mg/dL for an HbA1c of 6.9–7.1%. The root mean square of the errors between the actual mean sensor glucose concentration versus the value calculated using the regression equation was 14.3 mg/dL, whereas the median absolute difference was 10.1 mg/dL.
CONCLUSIONS
There is substantial individual variability between the measured versus calculated mean glucose concentrations. Consequently, estimated average glucose concentrations calculated from measured HbA1c values should be used with caution.
Publisher
American Diabetes Association
Subject
Advanced and Specialized Nursing,Endocrinology, Diabetes and Metabolism,Internal Medicine
Cited by
52 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献