Affiliation:
1. Oregon Health and Science University, Department of Medicine, Division of Endocrinology, Portland, Oregon;
2. Legacy Health, Division of Research, Portland, Oregon.
Abstract
OBJECTIVE
To minimize hypoglycemia in subjects with type 1 diabetes by automated glucagon delivery in a closed-loop insulin delivery system.
RESEARCH DESIGN AND METHODS
Adult subjects with type 1 diabetes underwent one closed-loop study with insulin plus placebo and one study with insulin plus glucagon, given at times of impending hypoglycemia. Seven subjects received glucagon using high-gain parameters, and six subjects received glucagon in a more prolonged manner using low-gain parameters. Blood glucose levels were measured every 10 min and insulin and glucagon infusions were adjusted every 5 min. All subjects received a portion of their usual premeal insulin after meal announcement.
RESULTS
Automated glucagon plus insulin delivery, compared with placebo plus insulin, significantly reduced time spent in the hypoglycemic range (15 ± 6 vs. 40 ± 10 min/day, P = 0.04). Compared with placebo, high-gain glucagon delivery reduced the frequency of hypoglycemic events (1.0 ± 0.6 vs. 2.1 ± 0.6 events/day, P = 0.01) and the need for carbohydrate treatment (1.4 ± 0.8 vs. 4.0 ± 1.4 treatments/day, P = 0.01). Glucagon given with low-gain parameters did not significantly reduce hypoglycemic event frequency (P = NS) but did reduce frequency of carbohydrate treatment (P = 0.05).
CONCLUSIONS
During closed-loop treatment in subjects with type 1 diabetes, high-gain pulses of glucagon decreased the frequency of hypoglycemia. Larger and longer-term studies will be required to assess the effect of ongoing glucagon treatment on overall glycemic control.
Publisher
American Diabetes Association
Subject
Advanced and Specialized Nursing,Endocrinology, Diabetes and Metabolism,Internal Medicine
Cited by
197 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献