Accuracy of an Electrochemical Sensor for Measuring Capillary Blood Ketones by Fingerstick Samples During Metabolic Deterioration After Continuous Subcutaneous Insulin Infusion Interruption in Type 1 Diabetic Patients

Author:

Guerci Bruno1,Benichou Muriel1,Floriot Michèle1,Bohme Philip1,Fougnot Sebastien1,Franck Patricia2,Drouin Pierre1

Affiliation:

1. Service de Diabétologie, Maladies Métaboliques & Maladies de la Nutrition, CIC-INSERM, Hôpital Jeanne d’Arc, Nancy, France

2. Laboratoire de Biochimie, Hôpital Central, Nancy, France

Abstract

OBJECTIVE—This study was designed to test the accuracy of capillary ketonemia for diagnosis of ketosis after interruption of insulin infusion. RESEARCH DESIGN AND METHODS—A total of 18 patients with type 1 diabetes treated by external pump were studied during pump stop for 5 h. Plasma and capillary ketonemia and ketonuria were determined every hour from 7:00 a.m. (time 0 min = T0) to 12:00 p.m. (time 300 min = T300). Plasma β-hydroxybutyrate (β-OHB) levels were measured by an enzymatic end point spectrophotometric method, and capillary β-OHB levels were measured by an electrochemical method (MediSense Optium meter). Ketonuria was measured by a semiquantitative test (Ketodiastix). Positive ketosis was defined by a value of ≥0.5 mmol/l for ketonemia and ≥4 mmol/l (moderate) for ketonuria. RESULTS—After stopping the pump, concentrations of β-OHB in both plasma and capillary blood increased significantly at time 60 min (T60) compared with T0 (P < 0.001), reaching maximum levels at T300 (1.30 ± 0.49 and 1.23 ± 0.78 mmol/l, respectively). Plasma and capillary β-OHB values were highly correlated (r = 0.94, P < 0.0001). For diagnosis of ketosis, capillary ketonemia has a higher sensitivity and negative predictive value (80.4 and 82.5%, respectively) than ketonuria (63 and 71.8%, respectively). For plasma glucose levels ≥250 mg/dl, plasma and capillary ketonemia were found to be more frequently positive (85 and 78%, respectively) than ketonuria (59%) (P = 0.017). The time delay to diagnosis of ketosis was significantly higher for ketonuria than for plasma ketonemia (212 ± 67 vs. 140 ± 54 min, P = 0.0023), whereas no difference was noted between plasma and capillary ketonemia. CONCLUSIONS—The frequency of screening for ketosis and the efficiency of detection of ketosis definitely may be improved by the use of capillary blood ketone determination in clinical practice.

Publisher

American Diabetes Association

Subject

Advanced and Specialized Nursing,Endocrinology, Diabetes and Metabolism,Internal Medicine

Cited by 102 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3