Abstract
An implantable glucose sensor is needed before a reliable artificial pancreas can be realized. The principles and current status of one such device, the glucose enzyme electrode, is presented and discussed. While monitoring glucose this enzyme sensor consumes enough oxygen to become oxygen-limited. This problem has been solved by developing hydrophobic membranes that are more permeable to oxygen than to glucose. Two types of membranes with this property made from (1) cross-linked albumin and (2) sebacyl chloride (nylon) are described. Placing these membranes over the glucose enzyme electrode solves the problem of oxygen limitation. Furthermore, the addition of this type of membrane increases the linear response range of the electrode to glucose to include the entire clinical range of interest (0-400 mg/dl). Other problems in developing an implantable glucose sensor are discussed. Competing strategies to achieve an implantable artificial pancreas without using electronic or mechanical components are presented and evaluated.