Rapid HbA1c Testing in a Community Setting

Author:

Carter Janette S12,Houston Caron A3,Gilliland Susan S1,Perez Georgia E1,Owen Charles L1,Pathak Dorothy R4,Little Randie R5

Affiliation:

1. Department of Medicine Albuquerque, New Mexico

2. University of New Mexico School of Medicine and the Veterans Administration Albuquerque, New Mexico

3. All Saints Senior Health Center Haltom City, Texas

4. Michigan State University Hast Lansing, Michigan

5. Departments of Pathology and Anatomical Sciences and Child Health, University of Missouri Columbia, Missouri

Abstract

OBJECTIVE To determine whether the DCA 2000 analyzer provides valid and reliable HbA1c results when used under field conditions and operated by nonmedical personnel. This study was part of a community diabetes education program, the Native American Diabetes Project, in which HbA1c was measured as an indicator of average glycemic control. RESEARCH DESIGN AND METHODS Two study samples were taken, the first in the spring of 1994 and the second in the spring of 1995. Seven community members in 1994 and six new community members in 1995 were trained over 2 days, using standard protocol, to operate the DCA 2000 HbA1c analyzer and to collect two capillary blood samples from participants in the Native American Diabetes Project. Duplicate DCA 2000 HbA1c measurements performed by the community workers were compared with measurements from a high-performance liquid chromatography (HPLC) system. Validity and reliability measures were calculated. RESULTS Of the participants, 43 were studied in 1994 and 14 in 1995. Comparison of the mean DCA 2000 results with those of HPLC showed high validity, with the absolute relative difference between the mean DCA 2000 and the external reference of HPLC (│ mean DCA 2000-HPLC │ /HPLC) as 4.0 and 2.0% for 1994 and 1995, respectively. The Pearson correlation coefficients (r) between these two measures were 0.968 and 0.996 for 1994 and 1995, respectively. While the 1994 data appeared to have less validity for values > 10%, they included only one value with a 60-min warm-up of the DCA analyzer. The 1995 data, all collected after a 60-min warm-up, had good correlation throughout the range of values. The within-run reliability was excellent, with an intraclass correlation coefficient of reliability of 0.959 and 0.975 for paired samples, for 1994 and 1995 respectively. The mean coefficient of variation for these paired measures was 3.0% in 1994 and 2.8% in 1995. Both validity and reliability were improved by changing the warm-up period of the DCA 2000 analyzer from 5 to 60 min. All correlation coefficients were statistically significant (P < 0.0001). CONCLUSIONS The DCA 2000 gave valid and reliable HbA1c results when operated in a community setting by nonmedical personnel. Extending the warm-up period of the device to 60 min slightly improved the validity and reliability of the test.

Publisher

American Diabetes Association

Subject

Advanced and Specialized Nursing,Endocrinology, Diabetes and Metabolism,Internal Medicine

Cited by 29 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3