Affiliation:
1. Departments of Endocrinology, Charing Cross and Chemical PathologyLondon, England
2. Chemical Pathology, Charing Cross and Chemical Pathology London, England
Abstract
A new reflectance blood glucose meter designed for hospital use, the Reflocheck (Boehringer-Mannheim, West Germany), was evaluated. The instrument showed excellent correlationwith a routine laboratory method (r = 0.996), and the coefficients of variation at high, medium, and low glucose levels were less than 4%.
Publisher
American Diabetes Association
Subject
Advanced and Specialized Nursing,Endocrinology, Diabetes and Metabolism,Internal Medicine
Cited by
11 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献