Abstract
OBJECTIVE
This study was designed to evaluate three different enzymatic methods for glucose measurement in plasma samples with special emphasis on glucose concentrations in the hypoglycemic range.
RESEARCH DESIGN AND METHODS
Glucose dehydrogenase (Hemo- Cue analyzer), glucose oxidase (YSI analyzer), and hexokinase (Abbott analyzer) methods were used to measure plasma samples that were obtained during research studies.
RESULTS
Mean glucose concentrations (n = 240) were 5.3 ± 0.2, 5.4 ± 0.2, and 5.6 ± 0.2 mM (95.6 ± 3.9, 96.7 ± 3.9, and 101.6 ± 4.0 mg/dl) using glucose dehydrogenase, glucose oxidase, and hexokinase, respectively (NS). In the hypoglycemic range, mean glucose concentrations with each method retained the same hierarchy of measurements: 2.7 ± 0.05, 2.8 ± 0.04, and 2.9 ± 0.03 mM (48.4 ± 0.9, 50.6 ± 0.8, and 52.3 ± 0.6 mg/dl) by glucose dehydrogenase, glucose oxidase, and hexokinase, respectively (P < 0.005). Individual glucose dehydrogenase measurements (n = 240) correlated well with glucose oxidase and hexokinase, r = 0.99, and were considerably easier to perform at the bedside. The differences between the glucose measurement methods were consistent and similar in low, normal, and high concentration ranges.
CONCLUSIONS
We conclude that any interpretation or comparison of critical clinical and research measurements of glucose in different settings take into account methodological differences, particularly in the hypoglycemic range.
Publisher
American Diabetes Association
Subject
Advanced and Specialized Nursing,Endocrinology, Diabetes and Metabolism,Internal Medicine
Cited by
34 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献