Use of an Automated Device for Alternative Site Blood Glucose Monitoring

Author:

Fineberg S. Edwin1,Bergenstal Richard M.2,Bernstein Robert M.3,Laffel Lori M.4,Schwartz Sherwyn L.5

Affiliation:

1. Indiana University School of Medicine, Indianapolis, Indiana

2. International Diabetes Center, Minneapolis, Minnesota

3. Regional Endocrinology Associates, Santa Fe, New Mexico

4. Joslin Diabetes Center, Boston, Massachusetts

5. Diabetes & Glandular Disease Clinic, San Antonio, Texas

Abstract

OBJECTIVE—To evaluate the accuracy, comfort, and ease of use of a new automated device for blood glucose monitoring using the arm as an alternative sampling site. RESEARCH DESIGN AND METHODS—These studies use an automated hand-held device that applies a small vacuum, lances the skin, transfers blood onto an electrochemical test strip, and measures glucose. Patients who had type 1 or type 2 diabetes and had received no prior training using this device were recruited from five diabetes clinics. Testing was performed by the patients using this device and by trained healthcare professionals. Blood glucose was measured by 354 patients: from the arm using the device, from the finger using a laboratory reference instrument, and from the finger using the device via the secondary test port. Each patient completed a questionnaire rating the level of pain and ease of use of the device. RESULTS—Blood glucose results in samples obtained from the arm with the automated device agreed well with finger-stick plasma glucose results using a reference instrument (regression slope 0.98, intercept 0.01 mmol/l [0.1 mg/dl], r = 0.96). Error grid analysis showed that 100% of the measurements fell within zones A and B. In the survey, 60% of the patients reported that arm testing with the automated device was “painless;” another 31% of the patients stated that it was “much less painful,” and 6% of patients considered using the device “less painful” than finger-stick testing. In a survey containing 15 questions for rating the ease of use with a scale of 1 to 6, the overall mean rating was 5.5. CONCLUSIONS—The automated device is easy to use and provides accurate glucose results; 97% of the patients found it less painful than finger-stick testing.

Publisher

American Diabetes Association

Subject

Advanced and Specialized Nursing,Endocrinology, Diabetes and Metabolism,Internal Medicine

Reference11 articles.

Cited by 51 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3