A U-Shaped Bioartificial Pancreas with Rapid Glucose-Insulin Kinetics: In Vitro Evaluation and Kinetic Modelling

Author:

Reach Gérard1,Jaffrin Michel Y1,Desjeux Jehan-François1

Affiliation:

1. Unité de Recherches sur le Diabète et la Nutrition chez l'Enfant INSERM U83, Paris Département de Biohydraulique et Instrumentation Médicale CNRS ERA 999 Université de Technologie Compiègne, France

Abstract

The lag in insulin release in response to glucose is an obstacle to the development of hybrid pancreatic devices, in which an artificial membrane protects transplanted islets against immune rejection. We designed a U-shaped bioartificial pancreas, in which the blood channel surrounds the islet chamber consisting of two flat membranes; blood circulates successively above the first membrane and then in the reverse direction, below the second membrane. Isolated rat islets were introduced into the chamber, which was perfused with Krebs buffer, and the kinetics of insulin release in response to glucose was determined. During a 20-min, 2.8–20-mM, square-wave glucose stimulation, insulin release in the effluent of the device rose from 0.7 ± 0.2 to 3.2 ±1.0 ng/100 islets/min (P < 0.05) within 3 min, and reached a maximal level of 12.8 ± 3.3 ng/100 islets/min at 10 min; 5 min after the return of the glucose concentration to substimulatory level, insulin release dropped from 11.3 ± 1.5 to 8.0 ± 1.7 ng/100 islets/min (P < 0.05), and reached basal value (1.0 ± 0.2 ng/100 islets/min) 40 min after the end of the stimulation. A 0.1-mM/L/min ramp increase in glucose concentration triggered a significant rise in insulin release (P < 0.02) when the glucose concentration reached 5.3 ± 0.2 mM; islets concomitantly perifused within a chamber set up without membrane responded to the same glucose stimulation 5 min earlier. For up to 1000 islets, insulin release in response to glucose was linearly correlated to the number of islets (N = 12, P < 0.01), indicating that insulin did not significantly inhibit its own secretion in this system. Finally, during glucose stimulation, the insulin concentration in the effluent from the chamber was found to be four times the concentration present at the turning point of the blood channel, suggesting that insulin was transferred into the perfusing medium in part by a countercurrent flux of ultrafiltrate crossing the membranes. We present herein the kinetic modelling of glucose and insulin transfer in this “Ultrafiltration chamber”, whose functional characteristics are compatible with closed-loop insulin delivery.

Publisher

American Diabetes Association

Subject

Endocrinology, Diabetes and Metabolism,Internal Medicine

Cited by 27 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3