Flow Mixing Enhancement from Balloon Pulsations in an Intravenous Oxygenator

Author:

Guzmán Amador M.1,Escobar Rodrigo A.2,Amon Cristina H.2

Affiliation:

1. Departamento de Ingeniería MecánicaUniversidad de Santiago de ChileAlameda 3363, Estación Central Santiago, Chile

2. Institute for Complex Engineered Systems, Departments of Mechanical Engineering and Biomedical Engineering, Carnegie Mellon University, 5000 Forbes Ave, Pittsburgh, PA 15213

Abstract

Computational investigations of flow mixing and oxygen transfer characteristics in an intravenous membrane oxygenator (IMO) are performed by direct numerical simulations of the conservation of mass, momentum, and species equations. Three-dimensional computational models are developed to investigate flow-mixing and oxygen-transfer characteristics for stationary and pulsating balloons, using the spectral element method. For a stationary balloon, the effect of the fiber placement within the fiber bundle and the number of fiber rings is investigated. In a pulsating balloon, the flow mixing characteristics are determined and the oxygen transfer rate is evaluated. For a stationary balloon, numerical simulations show two well-defined flow patterns that depend on the region of the IMO device. Successive increases of the Reynolds number raise the longitudinal velocity without creating secondary flow. This characteristic is not affected by staggered or non-staggered fiber placement within the fiber bundle. For a pulsating balloon, the flow mixing is enhanced by generating a three-dimensional time-dependent flow characterized by oscillatory radial, pulsatile longitudinal, and both oscillatory and random tangential velocities. This three-dimensional flow increases the flow mixing due to an active time-dependent secondary flow, particularly around the fibers. Analytical models show the fiber bundle placement effect on the pressure gradient and flow pattern. The oxygen transport from the fiber surface to the mean flow is due to a dominant radial diffusion mechanism, for the stationary balloon. The oxygen transfer rate reaches an asymptotic behavior at relatively low Reynolds numbers. For a pulsating balloon, the time-dependent oxygen-concentration field resembles the oscillatory and wavy nature of the time-dependent flow. Sherwood number evaluations demonstrate that balloon pulsations enhance the oxygen transfer rate, even for smaller flow rates.

Publisher

ASME International

Subject

Physiology (medical),Biomedical Engineering

Reference48 articles.

1. Extracorporeal Life Support for Adult Cardiorespiratory Failure;Anderson;Surgery (St. Louis)

2. Ten Years of Use of Extracorporeal Membrane Oxygenation (ECMO) in the Treatment of Acute Respiratory Insufficiency (ARI);Gillie;ASAIO Trans.

3. Current Status of Extracorporeal Membrane Oxygenation for Severe Respiratory Failure;Ichiba;Artif. Organs

4. Extracorporeal Respiratory Gas Exchange: A Look Into the Future;Kalobow;ASAIO Trans.

5. Extracorporeal Life Support in Critical Care Medicine;Sinard;J. Crit. Care

Cited by 4 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献