Modifying a Tilting Disk Mechanical Heart Valve Design to Improve Closing Dynamics

Author:

Herbertson Luke H.1,Deutsch Steven1,Manning Keefe B.1

Affiliation:

1. Department of Bioengineering, The Pennsylvania State University, University Park, PA 16802

Abstract

The closing behavior of mechanical heart valves is dependent on the design of the valve and its housing, the valve composition, and the environment in which the valve is placed. One innovative approach for improving the closure dynamics of tilting disk valves is introduced here. We transformed a normal Delrin occluder into one containing a ”dynamic liquid core” to resist acceleration and reduce the moment of inertia, closing velocity, and impact forces of the valve during closure. The modified occluder was studied in the mitral position of a simulation chamber under the physiologic and elevated closing conditions of 2500 mm Hg/s and 4500 mm Hg/s, respectively. Cavitation energy, detected as high-frequency pressure transients with a hydrophone, was the measure used to compare the modified valve with its unaltered counterpart. The cavitation potential of tilting disk valves is indicative of the extent of blood damage occurring during valve closure. Initial findings suggest that changes to the structure or physical properties of well established mechanical valves, such as the one described here, can reduce closure induced hemolysis by minimizing cavitation. Compared with a normal valve, the cavitation intensity associated with our modified valve was reduced by more than 66% at the higher load. Furthermore, the modified valve took longer to completely close than did the standard tilting disk valve, indicating a dampened impact and rebound of the occluder with its housing.

Publisher

ASME International

Subject

Physiology (medical),Biomedical Engineering

Reference22 articles.

1. Relative Blood Damage in the Three Phases of a Prosthetic Heart Valve Flow Cycle;Lamson;ASAIO J.

2. Cavitation

3. Johansen, P. , 2004, “Mechanical Heart Valve Cavitation,” Expert Review of Medical Devices, 1(1), pp. 95–104. 1743-4440

4. Cavitation Damage of Pyrolytic Carbon in Mechanical Heart Valves;Kafesjian;J. Heart Valve Dis.

5. High-Frequency Pressure Fluctuations Measured in Heart Valve Patients;Paulsen;J. Heart Valve Dis.

Cited by 4 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

1. Transvalvular flow;Principles of Heart Valve Engineering;2019

2. Measurement Technologies for Heart Valve Function;Advances in Heart Valve Biomechanics;2018

3. Effect of Cardiac Devices and Surgery on Vortex Formation;Vortex Formation in the Cardiovascular System;2011-11-11

4. Near Valve Flows and Potential Blood Damage During Closure of a Bileaflet Mechanical Heart Valve;Journal of Biomechanical Engineering;2011-09-01