Affiliation:
1. Department of Mechanical Engineering and Applied Mechanics, The University of Michigan, Ann Arbor, MI 48109-2125
Abstract
A precision multi-degree-of-freedom measuring (MDFM) system has been developed and implemented for the simultaneous measurement of straightness, pitch, yaw, and roll errors of the moving axes of a CMM. The system is based on the principles of laser alignment and autocollimator. Its measurement principles and the influence of laser beam drifts on its measurement quality have been investigated and some improvement schemes have been implemented. Through the measurements of actual as well as artificially created geometric errors of the CMM, it has been found that the system’s accuracy of measuring straightness error components is better than 1 μm and its accuracy for angular error measurements is better than 0.5 arcsec.
Cited by
49 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献