On the Importance of Modeling Stent Procedure for Predicting Arterial Mechanics

Author:

Zhao Shijia1,Gu Linxia2,Froemming Stacey R.3

Affiliation:

1. Department of Mechanical and Materials Engineering, University of Nebraska-Lincoln, Lincoln, NE 68588-0656

2. Department of Mechanical and Materials Engineering, University of Nebraska-Lincoln, Lincoln, NE 68588-0656 Nebraska Center for Materials and Nanoscience, Lincoln, NE 68588-0656 e-mail:

3. Hybrid Catheterization and Electrophysiology Laboratory, Children's Hospital and Medical Center, Omaha, NE 68114-4133

Abstract

The stent-artery interactions have been increasingly studied using the finite element method for better understanding of the biomechanical environment changes on the artery and its implications. However, the deployment of balloon-expandable stents was generally simplified without considering the balloon-stent interactions, the initial crimping process of the stent, its overexpansion routinely used in the clinical practice, or its recoil process. In this work, the stenting procedure was mimicked by incorporating all the above-mentioned simplifications. The impact of various simplifications on the stent-induced arterial stresses was systematically investigated. The plastic strain history of stent and its resulted geometrical variations, as well as arterial mechanics were quantified and compared. Results showed the model without considering the stent crimping process underestimating the minimum stent diameter by 17.2%, and overestimating the maximum radial recoil by 144%. It was also suggested that overexpansion resulted in a larger stent diameter, but a greater radial recoil ratio and larger intimal area with high stress were also obtained along with the increase in degree of overexpansion.

Publisher

ASME International

Subject

Physiology (medical),Biomedical Engineering

Reference32 articles.

1. The Stent Decade: 1987 to 1997. Stanford Stent Summit faculty;Am. Heart J.,1998

2. Endovascular Stent Design Dictates Experimental Restenosis and Thrombosis;Circulation,1995

3. The Relation Between the Arterial Stress and Restenosis Rate After Coronary Stenting;J. Med. Devices,2010

4. Increased Artery Wall Stress Post-Stenting Leads to Greater Intimal Thickening;Lab. Invest.,2011

5. Stainless and Shape Memory Alloy Coronary Stents: A Computational Study on the Interaction With the Vascular Wall;Biomech. Model Mechanobiol.,2004