Biomechanical Analysis of a New Carbon Fiber/Flax/Epoxy Bone Fracture Plate Shows Less Stress Shielding Compared to a Standard Clinical Metal Plate

Author:

Bagheri Zahra S.1,Tavakkoli Avval Pouria1,Bougherara Habiba2,Aziz Mina S. R.3,Schemitsch Emil H.345,Zdero Radovan15

Affiliation:

1. Department of Mechanical and Industrial Engineering, Ryerson University, Toronto, ON M5B-2K3, Canada

2. Department of Mechanical and Industrial Engineering (Eric Palin Hall), Ryerson University, 350 Victoria Street, Toronto, ON M5B-2K3, Canada e-mail:

3. Institute of Medical Science, University of Toronto, Toronto, ON M5S-1A8, Canada

4. Faculty of Medicine, University of Toronto, Toronto, ON M5S-1A8, Canada

5. Martin Orthopaedic Biomechanics Lab, St. Michael's Hospital, Toronto, ON M5B-1W8, Canada

Abstract

Femur fracture at the tip of a total hip replacement (THR), commonly known as Vancouver B1 fracture, is mainly treated using rigid metallic bone plates which may result in “stress shielding” leading to bone resorption and implant loosening. To minimize stress shielding, a new carbon fiber (CF)/Flax/Epoxy composite plate has been developed and biomechanically compared to a standard clinical metal plate. For fatigue tests, experiments were done using six artificial femurs cyclically loaded through the femoral head in axial compression for four stages: Stage 1 (intact), stage 2 (after THR insertion), stage 3 (after plate fixation of a simulated Vancouver B1 femoral midshaft fracture gap), and stage 4 (after fracture gap healing). For fracture fixation, one group was fitted with the new CF/Flax/Epoxy plate (n = 3), whereas another group was repaired with a standard clinical metal plate (Zimmer, Warsaw, IN) (n = 3). In addition to axial stiffness measurements, infrared thermography technique was used to capture the femur and plate surface stresses during the testing. Moreover, finite element analysis (FEA) was performed to evaluate the composite plate's axial stiffness and surface stress field. Experimental results showed that the CF/Flax/Epoxy plated femur had comparable axial stiffness (fractured = 645 ± 67 N/mm; healed = 1731 ± 109 N/mm) to the metal-plated femur (fractured = 658 ± 69 N/mm; healed = 1751 ± 39 N/mm) (p = 1.00). However, the bone beneath the CF/Flax/Epoxy plate was the only area that had a significantly higher average surface stress (fractured = 2.10 ± 0.66 MPa; healed = 1.89 ± 0.39 MPa) compared to bone beneath the metal plate (fractured = 1.18 ± 0.93 MPa; healed = 0.71 ± 0.24 MPa) (p < 0.05). FEA bone surface stresses yielded peak of 13 MPa at distal epiphysis (stage 1), 16 MPa at distal epiphysis (stage 2), 85 MPa for composite and 129 MPa for metal-plated femurs at the vicinity of nearest screw just proximal to fracture (stage 3), 21 MPa for composite and 24 MPa for metal-plated femurs at the vicinity of screw farthest away distally from fracture (stage 4). These results confirm that the new CF/Flax/Epoxy material could be a potential candidate for bone fracture plate applications as it can simultaneously provide similar mechanical stiffness and lower stress shielding (i.e., higher bone stress) compared to a standard clinical metal bone plate.

Publisher

ASME International

Subject

Physiology (medical),Biomedical Engineering

Reference75 articles.

1. Fractures of the Femur After Hip Replacement;Instructional Course Lectures,1995

2. Bone Ingrowth and Stress Shielding With a Porous Surface Coated Fracture Fixation Plate;J. Biomed. Mater. Res.,1979

3. The Influence of Different Degrees of Stiffness of Fixation Plates on Experimental Bone Healing;J. Orthop. Res.,1988

Cited by 72 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

1. Biomechanical design of a new proximal humerus fracture plate using alternative materials;International Journal for Numerical Methods in Biomedical Engineering;2024-09-07

2. A comprehensive review of the educational, clinical and rehabilitative applications of 3D printing technology in hand surgery;Plastic and Aesthetic Research;2024-08-29

3. Material Properties and Engineering Performance of Bone Fracture Plates Made from Plant Fiber Reinforced Composites: A Review;ACS Biomaterials Science & Engineering;2024-08-19

4. Fatigue performance impact of hybridization on Kevlar/flax and glass/flax hybrid composites;Proceedings of the Institution of Mechanical Engineers, Part L: Journal of Materials: Design and Applications;2024-07-26

5. Biomechanical Design Optimization of Clavicle Midshaft Fracture Plates: A Review;Journal of Medical and Biological Engineering;2024-06