This paper presents a t wo-dimensional 10 x 10 LED array system developed in-house at t he Standards and Calibration Laboratory (SCL) for the calibration of timing parameters of still image cameras and video cameras. In this paper, the circuit design and the calibration methods for cameras of different shutter types (rolling shutter or global shutter) are presented. This LED array may also be used to verify the shutter type of cameras. The frequency of the clock signal applied in the calibration is traceable to the SI through the cesium beam frequency standard maintained at SCL.
Cited by
2 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献
1. Calibration of Video Cameras Frame Rate based on Onsets Tracking Technique;2023 Joint Conference of the European Frequency and Time Forum and IEEE International Frequency Control Symposium (EFTF/IFCS);2023-05-15
2. Measuring timing properties of thermal infrared cameras;Measurement Science and Technology;2023-02-15