Augmentation of Dural Defect Repairs Strength With an Acrylic Plate in a Porcine Ex Vivo Model

Author:

Chen Philip G.1,Clampitt Michael R.2,Chorath Kevin T.1ORCID,Lin Ryan P.134,Weitzel Erik K.3,McMains Kevin C.3,Bunegin Leonid4

Affiliation:

1. Department of Otolaryngology—Head and Neck Surgery, University of Texas Health Science Center San Antonio, San Antonio, Texas

2. Department of Otorhinolaryngology, University of Oklahoma Health Sciences Center, Oklahoma City, Oklahoma

3. Uniformed Services University of Health Sciences and San Antonio Uniformed Health Sciences Educational Consortium, Joint Base, San Antonio, Texas

4. Department of Anesthesiology, University of Texas Health Science Center San Antonio, San Antonio, Texas

Abstract

Background Large skull base defects can be challenging to repair. This study uses a controlled ex vivo model to examine the failure pressures of various dural repairs of large skull base defects using mucosa with fibrin glue under 3 conditions: No Additional Support of the repair, support with a Foley catheter (Direct Support), and with Foley catheter contact over a rigid acrylic plate (Diffuse Support). Methods Failure pressures of dural repairs with and without support were determined in a porcine model using an ex vivo closed testing apparatus. In addition, 20 mm × 15 mm dural defects were created. Skull base repairs were performed using porcine dura as an underlay graft followed by a septal mucosa overlay. Saline was infused at 30 mL/h, applying even force to the underside of the graft until repair failure occurred for each condition (none, direct, and diffuse support). Five trials were performed per repair type for a total of 15 repairs. Results The mean failure pressures were as follows: No Additional Support, 6.494 ± 2.553 mm Hg; Direct Support, 5.103 ± 3.913 mm Hg; and Diffuse Support, 15.649 ± 2.638 mm Hg. A post hoc Bonferroni-Holm test demonstrated significant difference between No Additional Support and Diffuse Support ( P = .001), as well as Direct Support and Diffuse Support ( P = .002). Conclusion Support of dural repairs in this model withstood higher pressures when the Foley catheter’s support is distributed evenly using a flat acrylic plate. Use of this plate is the only repair tested in this model that tolerated normal adult supine intracranial pressures.

Publisher

SAGE Publications

Subject

General Medicine,Otorhinolaryngology,Immunology and Allergy

Cited by 2 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

1. Emerging Developments in Skull Base Reconstruction;Skull Base Reconstruction;2023

2. Something for Everyone in Rhinology and Allergy;American Journal of Rhinology & Allergy;2019-11

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3