Affiliation:
1. Papworth Hospital, Cambridge
Abstract
Vortex pumping has become popular for mechanical assist applications, but has only recently received serious consideration for use as the arterial pump of choice for routine open-heart surgery. We report on a prospective randomized study designed to evaluate this pump in routine use. Sixteen patients undergoing routine coronary artery surgery were randomized into two groups in which the only difference in equipment and technique was the use of a Stockert roller pump in group S and a Biomedicus vortex pump in group B. The groups were compared with respect to haematology, perioperative fluid balance, transfusion requirements, complement activation, haemolysis and microbubble transmission. The groups were reasonably well matched, with slightly longer mean bypass and ischaemic times in group S (94.8 vs 105.5 minutes and 75.7 and 83.8 minutes respectively). Group B patients showed significant improvements over group S patients with respect to preservation of platelet numbers, decreased complement activation and reduced microbubble transmission. The afterload sensitivity of the vortex pump did not present the perfusionist with any practical problems although a different technique is required for initiating and terminating bypass. We conclude that vortex pumping would seem to offer better blood handling for routine use. More extensive testing is required to establish whether or not this would be reflected in clinically measurable improved patient outcome.
Subject
Advanced and Specialised Nursing,Cardiology and Cardiovascular Medicine,Safety Research,Radiology Nuclear Medicine and imaging,General Medicine
Cited by
56 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献