New, optimized, dual-lumen cannula for veno-venous ECMO

Author:

von Segesser Ludwig K.1,Berdajs Denis2,Abdel-Sayed Saad1,Ferrari Enrico3,Halbe Maximilian4,Wilhelm Markus4,Maisano Francesco4

Affiliation:

1. Cardio-vascular Research, Department of Surgery and Anesthesiology, CHUV, Lausanne, Switzerland

2. Department of Cardiac Surgery, University Hospital Basel, Basel, Switzerland

3. Cardiocentro, Lugano, Switzerland

4. Department of Cardio-Vascular Surgery, University Hospital, Zurich, Switzerland

Abstract

Objective: The present study was designed to assess in vivo a new, optimized, virtually wall-less, dual-lumen, bi-caval cannula for veno-venous ECMO in comparison to a commercially available cannula. Methods: Veno-venous extracorporeal membrane oxygenation (ECMO) was carried out in a bovine study (n=5, bodyweight 75±5kg). Following systemic heparinization, ECMO was established in a trans-jugular fashion through a calibrated 23F orifice, using a new, optimized, virtually wall-less, dual-lumen, bi-caval 24F cannula (Smartcanula LLC, Lausanne, Switzerland) versus a commercially available 23F bi-caval, dual-lumen control cannula (Avalon Elite®, Maquet, Rastatt, Germany) in a veno-venous ECMO setup. Veno-venous ECMO was initiated at 500 revolutions per minute (RPM) and increased by incremental steps of 500 RPM up to 2500 RPM. Catheter outlet pressure, catheter inlet pressure, oxygen saturation and pump flow were recorded at each stage. Results: Mean flow accounted for 0.37±0.04 L/min for wall-less versus 0.29± 0.07 L/min for control at 500 RPM, 0.97±0.12 versus 0.67±0.06 at 1000 RPM, 1.60±0.14 versus 1.16±0.08 at 1500 RPM, 2.31±0.13 versus 1.52±0.13 for 2000 RPM and 3.02±0.5 versus 2.11±0.18 (p<0.004). The mean venous suction required was 19±8 mmHg for wall-less versus 20±3 mmHg for control at 500 RPM, 7±3 versus 9±4 for 1000 RPM, -11±10 versus -12±8 at 1500 RPM, -39±15 versus -49±10 for 2000 RPM and -60±28 versus -94±7 for 2500 RPM. The mean venous injection pressure accounted for 29±7 mmHg for wall-less versus 27±5 mmHg for control at 500 RPM, 50±6 versus 61±7 at 1000 RPM, 89±10 versus 99±17 for 1500 RPM, 142±14 versus 161±9 at 2000 RPM and 211±41 versus 252 ±3 for 2500 RPM. Conclusion: Compared to the commercially available control cannula, the new, optimized, virtually wall-less, dual-lumen, bi-caval 24F cannula allows for significantly higher blood flows, requires less suction and results in lower injection pressures in vivo.

Publisher

SAGE Publications

Subject

Advanced and Specialised Nursing,Cardiology and Cardiovascular Medicine,Safety Research,Radiology Nuclear Medicine and imaging,General Medicine

Cited by 5 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3