Comparison of image quality of head and neck lesions between 3D gradient echo sequences with compressed sensing and the multi-slice spin echo sequence

Author:

Kami Yukiko1ORCID,Chikui Toru1,Togao Osamu2,Ooga Masahiro3,Yoshiura Kazunori1

Affiliation:

1. Department of Oral and Maxillofacial Radiology, Faculty of Dental Science, Kyushu University, Fukuoka, Japan

2. Department of Clinical Radiology, Graduate School of Medical Sciences, Kyushu University, Fukuoka, Japan

3. Department of Medical Technology, Kyushu University Hospital, Fukuoka, Japan

Abstract

Background Although magnetic resonance imaging (MRI) provides excellent soft-tissue contrast, long acquisition times are major disadvantages. Purpose To evaluate the usefulness of compressed sensing (CS) for contrast-enhanced oral and maxillofacial MRI by comparing the 3D T1 turbo field echo with compressed SENSE (CS-3D-T1TFE) sequence with the multi-slice spin echo (MS-SE) sequence as the reference standard. Material and Methods Thirty patients with orofacial lesions participated in this study. The scan times for MS-SE and CS-3D-T1TFE were 5 min 56 s and 1 min 43 s, respectively. The signal-to-noise ratio (SNR) was calculated for quantitative analysis and seven parameters (degree of lesion conspicuity, motion artifacts, metal artifacts, pulsation artifacts, quality of fat suppression, homogeneity of blood vessel signal intensity, and overall image quality) were evaluated using a 5-point scale (5 = excellent, 1 = unacceptable) by two observers for qualitative analysis. For comparisons between MS-SE and CS-3D-T1TFE, the paired t-test was used. Results The SNR of CS-3D-T1TFE was higher than or equal to that of MS-SE. The CS-3D-T1TFE scores for motion artifacts, pulsation artifacts, and homogeneity of blood vessel signal intensity were higher than the corresponding MS-SE scores in assessments by both observers. The MS-SE scores for fat suppression were higher than or equal to the CS-3D-T1TFE scores. There were no significant differences in lesion conspicuity, metal artifacts, and overall image quality between the two sequences. Conclusion CS-3D-T1TFE imaging, less than 30% of the scan time for MS-SE, showed no image degradation while retaining equal or higher SNR and image quality.

Funder

JSPS KAKENHI

Publisher

SAGE Publications

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3