Comparison of Doppler ultrasonic and oscillometric devices (with or without proprietary optimisations) for non-invasive blood pressure measurement in conscious cats

Author:

Cerna Petra12ORCID,Archontakis Panos E3,Cheuk Hester OK4,Gunn-Moore Danièlle A15

Affiliation:

1. Department of Internal Medicine, Hospital for Small Animals, The Royal (Dick) School of Veterinary Studies, University of Edinburgh, Roslin, UK

2. The University of Veterinary and Pharmaceutical Sciences Brno, Czech Republic

3. Department of Clinical Science and Services, Royal Veterinary College, University of London, London, UK

4. First Vets, Newcastle upon Tyne, UK

5. The Roslin Institute, University of Edinburgh, Edinburgh, UK

Abstract

Objectives This study compared Doppler and oscillometric (PetMAP+) devices (with or without proprietary optimisations) for the non-invasive measurement of blood pressure in conscious cats. Methods Twenty-three cats were enrolled; however, five were excluded as fewer than five measurements were obtained for each assessment. All measurements were obtained according to American College of Veterinary Internal Medicine consensus guidelines. Oscillometric device modes A and B were operated according to the manufacturer’s guidelines. Doppler and oscillometric devices were used alternately as the first device. Results Systolic arterial blood pressure (SAP) measurements were obtained by Doppler (SAPd) and oscillometry; the mean of each set of five values was used for statistical analysis. There was a significant difference between SAPd and SAP measurements in oscillometric modes A ( P <0.001) and B ( P <0.001). While both modes measured SAP higher than SAPd, B had a smaller bias (+15.72 mmHg) and narrower limits of agreement (LOA). There was also a significant difference between SAPd and mean arterial pressure (MAP) on oscillometric modes A ( P = 0.002) and B ( P <0.001). Both modes’ MAP readings were lower than SAPd and oscillometric A MAP was closer to SAPd (–14.94 mmHg), with a smaller bias and narrower LOA. Conclusions and relevance The findings support that Doppler and oscillometric devices cannot be used interchangeably, with or without proprietary optimisations. Methodology should always be taken into account and reference intervals (RIs) need to be defined for the different methodologies. Until methodology-specific RIs are published, definitive diagnosis of hypertension and sub-staging of patients with kidney disease according to the International Renal Interest Society guidelines remains challenging.

Publisher

SAGE Publications

Subject

Small Animals