The performance of two heated water-bath humidifiers was examined in terms of airway temperatures and humidities. These were the simple heated tank [Marshall and Spalding, 1953) and the Fisher and Paykel No. 328 (Spence and Melville, 1972), both of which had the potential to deliver gases at temperatures above 40°C, and with a moisture content in excess of 50 mg/litre. The major advantage of the No. 328 was its heating wire in the delivery tube. This lessened or prevented condensation within the tube, and allowed the water bath to be maintained at lower temperatures. Independent heat controls for the water bath and the delivery tubing allow precision of humidities in the gases provided for inspiration through an artificial airway. When condensation is prevented, the temperature of the saturated gases leaving the humidifier limits the absolute humidity supplied. Monitoring and adjustment of airway temperature of the patient then allows selection of temperature and relative humidity in the inspired gas.
Subject
Anesthesiology and Pain Medicine,Critical Care and Intensive Care Medicine
Cited by
1 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献
1. An accurate method of measuring medical humidifier output;Clinical Physics and Physiological Measurement;1983-05