In vitro evaluation of how the presence of the stent retriever and microcatheter influences aspiration parameters in thrombectomy according to their position inside the aspiration catheter

Author:

Bernava Gianmarco1ORCID,Brina Olivier1,Reymond Philippe1,Rosi Andrea1,Hofmeister Jeremy1,Yilmaz Hasan1,Muster Michel1,Kulcsar Zsolt2,Lovblad Karl-Olof1,Machi Paolo1

Affiliation:

1. Division of Neuroradiology, Geneva University Hospitals, Geneva, Switzerland

2. Division of Neuroradiology, University Hospital of Zurich, Zurich, Switzerland

Abstract

Background Several variations of the combined thrombectomy technique for acute ischemic stroke using a stent retriever and aspiration catheter have been described. The aim of our study was to assess how the presence of the microcatheter and stent retriever affect the basic aspiration parameters, namely, flow rate and aspiration force, depending on their position within the aspiration catheter. Methods Two experimental set-ups were designed to assess changes in flow rate and aspiration force according to the position of the stent retriever and microcatheter within the aspiration catheter. Results The transition of the stent retriever and microcatheter from the distal to proximal position resulted in a progressive increase in the flow rate, but with no impact on aspiration force. Additionally, the size of the stent retriever had no significant effect on flow rate changes and the reduction in flow rate was related to the microcatheter diameter. Negative pressure generated inside the aspiration catheter impacted on its distal segment located beyond the radiopaque marker, thus leading to its partial collapse. As a consequence, the measured aspiration force was lower than the theoretical aspiration force level for all tested aspiration catheters. Conclusions In our experimental model, the position of the stent retriever and microcatheter within the aspirator catheter affected the flow rate, but not the aspiration force. Negative pressure generated within the aspiration catheter appeared to determine a partial collapse of the distal segment that resulted in a less effective aspiration force than the theoretical aspiration force level.

Funder

Schweizerischer Nationalfonds zur Förderung der Wissenschaftlichen Forschung

Publisher

SAGE Publications

Subject

Immunology

Cited by 2 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3