Knowledge-based iterative reconstructions for imaging of coronary artery stents: first in-vitro experience and comparison of different radiation dose levels and kernel settings

Author:

Hickethier Tilman1ORCID,Baeßler Bettina1,Kroeger Jan Robert1,Müller Dirk2,Maintz David1,Michels Guido3,Bunck Alexander C1

Affiliation:

1. Department of Radiology, University Hospital of Cologne, Cologne, Germany

2. Clinical Science CT, Philips Germany GmbH, Hamburg, Germany

3. Department III of Internal Medicine, University Hospital of Cologne, Cologne, Germany

Abstract

Background Advanced knowledge-based iterative model reconstructions (IMR) became recently available for routine computed tomography (CT). Using more realistic physical models it promises improved image quality and potential radiation dose reductions, both possibly beneficial for non-invasive assessment of coronary stents. Purpose To evaluate the influence of different IMR settings at different radiation doses on stent lumen visualization in comparison to filtered back projection (FBP) and first-generation (hybrid) iterative reconstruction (HIR). Material and Methods Ten coronary stents in a coronary phantom were examined at four different dose settings (120 kV/125 mAs, 120 kV/75 mAs, 100 kV/125 mAs, 100 kV/75 mAs). Images were reconstructed with stent-specific FBP and HIR kernels and with IMR using CardiacRoutine (CR) and CardiacSharp (CS) settings at three different iteration levels. Image quality was evaluated using established parameters: image noise; in-stent attenuation difference; and visible lumen diameter. Results Image noise was significantly lower in IMR than in corresponding HIR and FBP images. At lower radiation doses, image noise increased significantly except with IMR CR3 and IMR CS3. Visible lumen diameters were significantly larger with IMR CS than with FBP, HIR, and IMR CR. IMR CR showed the smallest attenuation difference, while attenuation was artificially decreased extensively with IMR CS. FBP and HIR showed moderately increased in-stent attenuations. No relevant influence of used radiation doses on visible lumen diameters or attenuation differences was found. Conclusion IMR CR reduces image noise significantly while offering comparable stent-specific image quality in comparison to FBP and HIR and therefore potentially facilitates stent lumen delineation. Utilization of IMR CS for stent evaluation seems unfavorable due to artificial image alterations.

Publisher

SAGE Publications

Subject

Radiology, Nuclear Medicine and imaging,General Medicine,Radiological and Ultrasound Technology

Cited by 2 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3