Minor impact of constraint from perivascular flow probes on wave intensity analysis

Author:

Mynard Jonathan P1234ORCID,Smolich Joseph J12

Affiliation:

1. Heart Research, Clinical Sciences, Murdoch Children’s Research Institute, Parkville, VIC, Australia

2. Department of Paediatrics, University of Melbourne, Parkville, VIC, Australia

3. Department of Biomedical Engineering, University of Melbourne, Parkville, VIC, Australia

4. Department of Cardiology, Royal Children’s Hospital, Parkville, VIC, Australia

Abstract

Perivascular flow probes are considered the gold-standard for measuring volumetric blood flow in animal studies. Although flow probes are generally placed non-constrictively around the vessel of interest, pressure-elevating interventions performed during an experiment may lead to vessel expansion and some probe-vessel impingement, particularly in highly compliant vessels such as adult sheep aorta or major pulmonary arteries in fetus lambs. This study assessed to what extent such mild flow probe constraint may impact on wave intensity analysis. We also investigated whether errors arising from flow probe constraint could explain apparent pressure reflection indices ( Rp > 1) that have been observed in fetus lamb pulmonary arteries under some experimental conditions. These questions were investigated with one-dimensional models of an adult sheep aorta and fetus lamb pulmonary artery, with a virtual flow probe incorporated as a non-linear external constraint term in the vessel constitutive equation. Model-derived flow and pressure were subjected to standard analysis procedures that would be applied experimentally (correcting for apparent velocity lags and calculating wave speed via the PU-loop method). For the adult sheep model, simulations covering a wide range of haemodynamic conditions revealed a mostly minor effect (<10%) of probe constraint on the intensity and pressure effects of the three major waves (forward compression wave, forward decompression wave, backward compression wave). Moreover, flow probe constraint had essentially no impact on Rp in the fetus lamb model, suggesting that such constraint is unlikely to be responsible for an observed Rp > 1. Mild flow probe constraint is likely to have little impact on wave intensity analysis.

Funder

National Health and Medical Research Council

National Heart Foundation of Australia

Publisher

SAGE Publications

Subject

Mechanical Engineering,General Medicine

Cited by 1 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

1. Physiological fluid mechanics: A special Issue with a taster of forefront research;Proceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine;2020-10-10

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3