Abstract

Compression plates are widely used in orthopaedic surgeries for internal fixation of fractured femurs. To fix the plate and thus to provide compression to a fracture, the self-tapping bone screws are tightened through predrilled pilot holes of smaller diameter. Preliminary investigation showed that the holes drilled with the inappropriate cutting parameters cause mechanical and thermal damages to the local host bone, which further lead to loosening of internal fixations. In this paper, the mechanistic models to predict the thrust forces and torques during bone drilling were developed, using a 3.20 mm diameter drill bit. As a procedure, the cutting action was investigated at three different regions of the drill point, namely cutting lips, secondary cutting edges and indentation zone. The models employed the analytical approach to account for the drill-bit geometry and cutting parameters, and an empirical approach to account for the material and friction properties. To complete the procedure, calibration experiments were conducted on bovine cortical femurs with two different spindle speeds (1000 and 3000 r/min) and feeds (0.03 and 0.06 mm/rev), and then the specific normal and friction coefficients were determined. The developed mechanistic models were validated with different ranges of parameters (500–3500 r/min speeds, and 0.02–0.07 mm/rev feeds) those commonly involved in manual and robot-assisted surgery. The validation study revealed that the thrust forces predicted using the mechanistic models showed a maximum error of only 5.80%. However, the torques predicted from the mechanistic model found with more error than the thrust forces. The predominant reasons for this under-prediction might because of the extrapolation used to determine the specific cutting pressures, slip-line field applied to the indentation zone instead of compressive fracture, and chip clogging involved during the bone drilling as demonstrated in earlier studies. Despite the deviations, the developed mechanistic models satisfactorily follow the trends of the thrust forces and torques experienced during bone drilling. The outcomes can be used to practice the bone drilling procedure and monitor the effect of process parameters on thrust forces and torques in the in-silico environment before performing actual surgery.

Publisher

SAGE Publications

Subject

Industrial and Manufacturing Engineering,Mechanical Engineering

Cited by 6 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3