Detection of 5-fluorouracil surface contamination in near real time

Author:

Smith Jerome P1,Sammons Deborah L1,Pretty Jack R1,Kurtz Kristine S2,Robertson Shirley A1,DeBord D Gayle1,Connor Thomas H1,Snawder John E1

Affiliation:

1. Division of Applied Research and Technology, National institute for Occupational Safety & Heath, US Centers for Disease Control and Prevention, Cincinnati, OH, USA

2. Bureau Veritas North America, Novi, MI, USA

Abstract

Objectives Contamination of workplace surfaces by antineoplastic drugs presents an exposure risk for healthcare workers. Traditional instrumental methods to detect contamination such as gas chromatography-mass spectrometry (GC-MS) or liquid chromatography-tandem mass spectrometry (LC-MS/MS) are sensitive and accurate but expensive and incapable of producing results in real time. This limits their utility in preventing worker exposure. We are currently developing monitors based on lateral flow immunoassay that can detect drug contamination in near real time. In this report, we describe the laboratory performance of a 5-fluorouracil (5-FU) monitor. Methods The monitor was evaluated by spiking ceramic, vinyl, composite, stainless steel, and glass surfaces of 100 cm2 area with 5-FU masses of 0, 5, 10, 25, 50, and 100 ng. The surface was sampled with a wetted cotton swab, the swab was extracted with buffer, and the resulting solution was applied to a lateral flow monitor. Two ways of evaluating the response of these monitors were used: an electronic method where a lateral flow reader was used for measuring line intensities, and a visual method where the intensity of the test line was visually compared to the control line. Results The 5-FU monitor is capable of detecting 10 ng/100 cm2 (0.1 ng/cm2) using the electronic reader and 25 ng/100 cm2 (0.25 ng/cm2) using the visual comparison method for the surfaces studied. The response of the monitors was compared to LC-MS/MS results for the same samples for validation and there was good correlation of the two methods but some differences in absolute response, especially at higher spiking levels for the surface samples.

Publisher

SAGE Publications

Subject

Pharmacology (medical),Oncology

Cited by 5 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3