Abstract
This study represents the first attempt to electrically pace the paralyzed human larynx. The goal was to determine if electrical stimulation of the posterior cricoarytenoid muscle could produce functional abduction of the vocal fold in pace with inspiration. An external apparatus was used to sense inspiration and reanimate the unilaterally paralyzed larynx of a thyroplasty patient. Stimuli were delivered through a needle electrode to locate and pace the abductor muscle. The magnitude of electrically induced abduction was comparable to spontaneous movement on the normal side. The abduction was appropriately timed with inspiration; this finding demonstrated that this simple pacing system could effectively modulate stimulation with patient respiration.
Subject
General Medicine,Otorhinolaryngology
Cited by
59 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献