Affiliation:
1. Aerodynamisches Institut, Technical University, Aachen - FRG
Abstract
The report describes the development of heart valve test standards. The aim is comprehensive quality assurance by in vitro tests. The project includes three test fields: general basis, development and definition of test methods and test devices and comparative in vitro assessment of valves for the definition of minimum requirements. A preliminary list of test parameters and test steps has been defined: geometrical, flow, deformation, force, and conditioning parameters. A system of geometrical parameters has been developed for standardized aortic models. Geometrical parameters of 31 valves of six types and different sizes underline a close correlation between geometrical and hemodynamic parameters. The relative ostium cross-section Ae/AT increases with valve size and lies between 0.3 and 0.5. Two new measurement devices with quasi-steady flow are proposed as quick testers for leakage flow and pressure loss.
Subject
Biomedical Engineering,Biomaterials,General Medicine,Medicine (miscellaneous),Bioengineering
Reference4 articles.
1. KöhlerJ.: Quality assurance of artificial heart valves — In vitro testing — Heart valve engineering; London: Mechanical Engineering Publications Ltd; 1986; 89–98.
Cited by
3 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献