Evaluation of Blood Cell Destruction by Measuring Occlusion Distance

Author:

Kato Shota1,Handa Tadashi1,Yoshioka Jun2,Nakadate Kazuhiko3,Nomura Yasutomo4,Kijima Hitoshi1

Affiliation:

1. Department of Medical Technology and Clinical Engineering, Gunma University of Health and Welfare, 191-1 Kawamagarimachi Maebashi Gunma 371-0823, JAPAN

2. Sendai Red Cross Hospital Medical Technology Department, Clinical Engineering Technology Division, 2-43-3, Hakugiyamahonchou Sendai Miyagi 982-8501, JAPAN

3. Department of Basic Science Educational and Research Center for Pharmacy, Meiji Pharmaceutical University, 2-522-1Nojiri Kiyose Tokyo 204-8588, JAPAN

4. Maebashi Institute of Technology Graduate School of Environment and Biotechnology, 460-1, Kamisadori Maebashi Gunma 371-0816, JAPAN

Abstract

Roller pumps are commonly used for electric motor-driven blood purification. Even the optimal occlusion for a roller pump is stimulated in JIS (Japanese Industrial Standard) -T1603, the blood cells can be destroyed if an applied pressure is too strong on the tube. On the other hand, the perfused blood volume might decrease if the pressure becomes weaker. Therefore, skilled operation is required. However, as there are no techniques to automatically measure occlusion, a highly reproducible method is urgently required to obtain an optimal setting. In this study, we classified the occlusion specified in JIS-T1603 into five categories (3, 6, 9, 12, and 15 drops/min) and measured those using a laser sensor. The distance between each occlusion was only a few microns. Based on the microscopic observation of the blood cell morphology at each occlusion, the blood cells with normal outlines were classified as normal blood cells, while those with protrusions were labeled as acanthocytes. Further, we calculated the normalized milligram index of hemolysis (mgNIH) to confirm hemolysis for each occlusion. By classifying occlusion into five categories and converting them into distances, we derived a safe, easy, and highly reproducible method.

Publisher

World Scientific and Engineering Academy and Society (WSEAS)

Subject

General Agricultural and Biological Sciences,General Biochemistry, Genetics and Molecular Biology,General Medicine,General Neuroscience

Cited by 1 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

1. Defining Elements of Roller Pump Occlusion in Cardiopulmonary Bypass Surgery;WSEAS TRANSACTIONS ON BIOLOGY AND BIOMEDICINE;2024-04-01