Affiliation:
1. Department of Pharmacokinetics, Kyoto Pharmaceutical University, Kyoto 607 8414, Japan
Abstract
Background: Pemafibrate, a novel selective peroxisome proliferator-activated receptor-α modulator, is prescribed for patients with dyslipidemia. To investigate other potential nonlipid-related effects of pemafibrate, the sensitive and rapid quantitation method for pemafibrate was required. Results: The developed LC–MS/MS assay method exhibited excellent accuracy, precision, sensitivity, stability, no matrix effect and high recovery. The LOQ (0.05 ng/ml) and run time (6.0 min) were superior to previous reports. The calibration curve showed good linearity over the wide concentration range (0.05–100.00 ng/ml). This validated method was successfully applied in a rat pharmacokinetic study using lower doses (0.02 or 0.10 mg/kg) than have been previously reported. Conclusion: This method can support gathering data for the evaluation of pemafibrate in future studies.
Subject
Medical Laboratory Technology,Clinical Biochemistry,General Pharmacology, Toxicology and Pharmaceutics,General Medicine,Analytical Chemistry