A comparison of bolus insulin dose errors based on results of a clinical trial of five blood glucose monitoring systems

Author:

Pardo Scott1,Dunne Nancy2,Simmons David A2

Affiliation:

1. Ascensia Diabetes Care, Valhalla, NY 10595, USA

2. Ascensia Diabetes Care, Parsippany, NJ 07054, USA

Abstract

Aim: Inaccurate blood glucose monitoring system (BGMS) results may lead to insulin dosing errors and adverse clinical outcomes. Results & methodology: This post-hoc analysis used a model to estimate the bolus insulin dose error associated with each of the five BGMSs, for a hypothetical person with diabetes (assuming a standardized meal and target blood glucose of 100 mg/dl). Differences in dose-error distribution between BGMSs were statistically tested. The 95% dose-error range for each BGMS was (insulin units): CONTOUR®PLUS, -1.1–0.7; Accu-Chek® Active, -2.4–0.7; Accu-Chek® Performa, -2.9–0.8; FreeStyle Freedom, from -5.5 to -0.5; OneTouch® SelectSimple, -4.1−3.0. Conclusion: The CONTOUR®PLUS BGMS was associated with a statistically significantly smaller model-estimated median bolus insulin dose-error and dosing error range, compared with the other BGMSs.

Publisher

Future Science Ltd

Subject

Pharmaceutical Science