End-Tidal Control Versus Manual Control of Inhalational Anesthesia Delivery: A Randomized Controlled Noninferiority Trial

Author:

McCabe Melissa D.1,Dear Guy de L.2,Klopman Matthew A.3,Garg Kritika4,Seering Melinda S.5

Affiliation:

1. Department of Anesthesiology, Loma Linda University School of Medicine, Loma Linda, California

2. Department of Anesthesiology and Pediatrics, Duke University Medical Center, Durham, North Carolina

3. Department of Anesthesiology, Emory University School of Medicine, Atlanta, Georgia

4. Biostatistics Division of Clinical Operations Department, GE HealthCare, Chicago, Illinois

5. Department of Anesthesiology, Roy and Lucille Carver College of Medicine, University of Iowa Health Care, Iowa City, Iowa.

Abstract

BACKGROUND: Precise anesthesia delivery helps ensure amnesia, analgesia, and immobility. Conventionally, the end-tidal anesthetic concentration is maintained through manual adjustment of the fresh gas flow and anesthetic vaporizer output. Some anesthesia delivery systems can deliver and maintain clinician-selected end-tidal anesthetic agent (EtAA) concentration using a modified closed-loop system. We evaluated the performance of an End-tidal Control (EtC) system on the Aisys CS2 anesthesia machine (GE HealthCare). We hypothesized EtC anesthetic delivery would be noninferior to manually controlled anesthetic delivery. METHODS: The Multi-site Anesthesia randomized controlled STudy of End-tidal control compared to conventional Results (MASTER) Trial evaluated anesthetic delivery in 210 adult patients receiving inhaled anesthesia. Patients were randomized to either EtC or manual control (MC) anesthetic delivery. The primary objective was to determine whether, compared to conventional anesthesia practice, EtC achieves and maintains clinician-specified EtAA and end-tidal oxygen (Eto 2) concentrations within defined noninferiority limits. Noninferiority was concluded if the lower limit of the 95% confidence interval (CI) of the difference between the percent duration within the acceptable range (5% of steady state or a margin of ~10% of each agent’s minimum alveolar concentration) for EtC and MC was ≥ −5% for both EtAA and Eto 2 . Secondary objectives included performance measures: response time: time required to attain 90% of the first desired EtAA, overshoot: amount the controller (or vaporizer delivery) exceeded the desired EtAA, and accuracy: average deviation from the desired EtAA. RESULTS: EtC achieved and sustained targeted EtAA and Eto 2 concentrations within the noninferiority threshold. The EtAA was within 5% of the desired value 98% ± 2.05% of the time with EtC compared to 45.7% ± 31.7% of the time with MC (difference 52.3% [95% CI, 45.9%–58.6%], P < .0001). For Eto 2, EtC was within the noninferiority limit 86.3% ± 22.8% of the time compared with MC at 41% ± 33.3% (P < .0001, difference 45.3% [95% CI, 36.1%–54.5%]). The median response time for achieving 90% of the initial EtAA desired value was 75 seconds with EtC and 158 seconds with MC (P = .0013). EtC exhibited a median overshoot of 6.64% of the selected EtAA concentration, whereas MC often failed to reach the clinician’s desired value. The difference in median percent deviation from desired EtAA value was 15.7% ([95% CI, 13.5%–19.0%], P < 0001). CONCLUSIONS: EtC achieves and maintains the EtAA and Eto 2 concentration in a manner that is noninferior to manually controlled anesthesia delivery.

Publisher

Ovid Technologies (Wolters Kluwer Health)

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3