Author:
Tognolo Lucrezia,Coraci Daniele,Farì Giacomo,Vallenari Valeria,Masiero Stefano
Abstract
Appendicular skeletal mass is commonly used to assess the loss in muscle mass and US represents a valid, and reliable method. However, the procedural protocols are still heterogeneous. The aim of this study was to compare the intertransducers validity of thickness, width, and CSA measurements of RF muscle. The AP, LL and CSA of RF muscle were evaluated with both linear and curve probes in ten healthy subjects and six sarcopenic patients. In the healthy group the mean AP diameters measured with the linear array were significantly higher than those measured with the curved array. AP and CSA were higher in the healthy group compared with the sarcopenic group with both transducers. There was a positive correlation between weight and LL diameter, and a negative correlation between age and muscle AP, measured with the linear probe. Both linear and curved probes represent valid methods in US evaluation of the CSA of the RF muscle. However, in the healthy subjects, the thickness and width of the of the same muscle, are affected by the type of probe.
Subject
Cell Biology,Neurology (clinical),Molecular Biology,Orthopedics and Sports Medicine
Cited by
6 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献