Affiliation:
1. All-Russia Scientific Research Institute of Aviation Materials
Abstract
Beryllium products exhibiting a low level of absorption of the radiation energy are widely used in scientific instrumentation design (x-ray technology, radiation detectors, etc.). We present the results of studying the leak tightness of products (disks, plates) made of technical sintered beryllium of standard purity and foil obtained by «warm» rolling from high-purity beryllium. The relevant standards and requirements for testing are given. The leak tightness control was performed using mass spectrometric helium leak detectors with forevacuum backing pumps (oil and dry diaphragm pumps) and specialized vacuum equipment. The parameters of tightness of samples made of technical sintered beryllium were determined. The level of the helium signal during blowing was (0.6 – 7.4) × 10–11 Pa · m3/sec, which corresponds to the tightness standard of foreign analogues and matches the requirements of domestic manufacturers of x-ray equipment. The data spread tended to increase due to the growth of the background value. The obtained results can be used to improve high-tech equipment intended for flaw detection, medical devices, rapid analysis of ore raw materials, radiation safety equipment, etc.