Average of Patient Deltas: Patient-Based Quality Control Utilizing the Mean Within-Patient Analyte Variation

Author:

Cembrowski George S12,Xu Qian3,Cervinski Mark A45

Affiliation:

1. Laboratory Medicine and Pathology, University of Alberta, Edmonton, AB, Canada

2. Cembrowski Cembrowski Quality Control Consulting, Edmonton, AB, Canada

3. University of Manitoba School of Medicine, Winnipeg, MB, Canada

4. Department of Pathology and Laboratory Medicine, Geisel School of Medicine at Dartmouth, Hanover, NH, USA

5. Department of Pathology and Laboratory Medicine, Dartmouth-Hitchcock Medical Center, Lebanon, NH, USA

Abstract

Abstract Background Because traditional QC is discontinuous, laboratories use additional strategies to detect systematic error. One strategy, the delta check, is best suited to detect large systematic error. The moving average (MA) monitors the mean patient analyte value but cannot equitably detect systematic error in skewed distributions. Our study combines delta check and MA to develop an average of deltas (AoD) strategy that monitors the mean delta of consecutive, intrapatient results. Methods Arrays of the differences (delta) between paired patient results collected within 20–28 h of each other were generated from historical data. AoD protocols were developed using a simulated annealing algorithm in MatLab (Mathworks) to select the number of patient delta values to average and truncation limits to eliminate large deltas. We simulated systematic error by adding bias to arrays for plasma albumin, alanine aminotransferase, alkaline phosphatase, amylase, aspartate aminotransferase, bicarbonate, bilirubin (total and direct), calcium, chloride, creatinine, lipase, sodium, phosphorus, potassium, total protein, and magnesium. The average number of deltas to detection (ANDED) was then calculated in response to induced systematic error. Results ANDED varied by combination of assay and AoD protocol. Errors in albumin, lipase, and total protein were detected with a mean of 6 delta pairs. The highest ANDED was calcium, with a positive 0.6-mg/dL shift detected with an ANDED of 75. However, a negative 0.6-mg/dL calcium shift was detected with an ANDED of 25. Conclusions AoD detects systematic error with relatively few paired patient samples and is a patient-based QC technique that will enhance error detection.

Publisher

Oxford University Press (OUP)

Subject

Biochemistry (medical),Clinical Biochemistry

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3