Abstract
Abstract
Pattern recognition techniques--discriminant analysis and principal component analysis--are utilized in selecting the wavelengths for monitoring, by reflectance spectroscopy, color-generating reactions involving uric acid and cholesterol in serum. The data base we used was accumulated by a rapid-scanning reflectance spectrophotometer that measured reflectance at 16 wavelengths every 5 s after the reaction was initiated. The data were then analyzed in multidimensional space mainframe computer with commercial statistical software packages. The most appropriate wavelengths were those that yielded the largest generalized distance between analyte concentration by discriminant analysis and the largest weighting coefficient by principal component analysis. For uric acid, taking the ratio of reflectance measured at two wavelengths instead of at a single wavelength much better separates the clinically significant concentrations. For cholesterol, the initiated. The data were then analyzed in multidimensional space hemoglobin, can be clearly demonstrated y the "pattern" generated with principal component analysis. generalized distance between analyte generalized distance between analyte concentration by discriminant
Publisher
Oxford University Press (OUP)
Subject
Biochemistry (medical),Clinical Biochemistry
Cited by
4 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献