Affiliation:
1. Laboratory Services, Division of Metpath Inc., San Jose, CA 95133
Abstract
Abstract
Hemolysates from 102 different patients previously assessed for the presence of hemoglobin variants by cellulose acetate electrophoresis were reanalyzed with the Diamat, a microprocessor-controlled step-gradient HPLC technique designed to determine glycohemoglobin (Hgb A1c). This pool contained 81 abnormal specimens, each with one of seven different variant abnormalities. In all cases the HPLC technique correctly detected the presence or absence of a variant. The common S trait gave a large peak immediately after Hgb A, and the SS and SC variants gave clearly abnormal patterns, caused in part by the absence of Hgb A. The pattern from EE variants was distinguished by the appearance of the glycated fraction Hgb E1c, which is eluted at 4.7 instead of 4.2 min, whereas the AE pattern had two glycated peaks. Hemoglobins F, "fast", S, and C were discerned by the presence of a major peak at 3, 5.2, 6.5, and 7.5 min, respectively. The findings suggest that the Diamat is a convenient tool to detect and help diagnose variants in a screening pool.
Publisher
Oxford University Press (OUP)
Subject
Biochemistry, medical,Clinical Biochemistry
Cited by
8 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献